| 雾度计计量校准的应用与思考 |
| The application and consideration of haze meter's measurement and calibration |
| 投稿时间:2014-02-18 |
| DOI:10.3969/j.issn.1005-5630.2014.05.008 |
| 中文关键词: 雾度计 校准 准确性 |
| 英文关键词:haze meter calibration accuracy |
| 基金项目: |
|
| 摘要点击次数: 3063 |
| 全文下载次数: 3618 |
| 中文摘要: |
| 针对雾度计校准和JJF 1303-2011《雾度计校准规范》使用过程中遇到的一些实际问题进行阐述.主要集中探讨环境因素对雾度计测量结果的影响、雾度标准片的示值误差、雾度仪生产过程中的漏洞3个方面,并给出了改进的建议,供行业内相关人员参考. |
| 英文摘要: |
| The practical problems are introduced which are encountered in the instrument calibration and JJF 1303-2011 ‘Calibration Specification for Haze meter’ in the practical using. The problems mainly focus on three parts:the influence of environmental factors over the measurement results, the indication error of haze standard plate, and the flaw of producing haze meter. Then the improved suggestions are given for the relevant reference. |